کالیبراتور های سری 725
(Fluke 725 Multifunction Process Calibrator)
img1

 

 

بررسی اجمالی

Fluke 725 Multifunction Process Calibrator

مشخصات فنی

Measurement Accuracy
Voltage DC 30.000 V 0.02% + 2 counts (upper display)
30.000 V 0.02% + 2 counts (lower display)
100.00 mV 0.02% + 2 counts
-10.00 mV to 75.00 mV 0.025% + 1 count (via TC connector)
Current DC 24.000 mA 0.02% + 2 counts
Resistance 0.0 to 400.0 Ω 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)
401 to 1500 Ω 0.5 Ω (4-wire), 1 Ω (2- and 3-wire)
1500 to 3200 Ω 1 Ω (4-wire), 1.5 Ω (2- and 3-wire)
Frequency 2.0 to 1000.0 CPM 0.05% + 1 count
1.0 to 1100.0 Hz 0.05% + 1 count
1.00 to 10.00 kHz 0.05% + 1 count
Sensitivity 1 V peak-to-peak-minimum
Pressure Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC 100.00 mV 0.02% + 2 counts
10.000 V 0.02% + 2 counts
-10.00 mV to 75.00 mV 0.025% + 1 count (via TC connector)
Current DC 24.000 mA (source) 0.02% + 2 counts
24.000 mA (simulate) 0.02% + 2 counts
Resistance 15.0 to 400.0 Ω 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
401 to 1500 Ω 0.5 Ω (excitation current 0.05 to 0.8 mA)
1500 to 3200 Ω 1 Ω (excitation current 0.05 to 0.4 mA)
Frequency 2.0 to 1000.0 CPM 0.05%
1.0 to 1100.0 Hz 0.05%
1.00 to 10.00 kHz 0.25%
Waveform 5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure accuracy NI-120 0.2°C
PT-100 (385) 0.33°C
PT-100 (393) 0.3°C
PT-100 (JIS) 0.3°C
PT-200 (385) 0.2°C
PT-500 (385) 0.3°C
PT-1000 (385) 0.2°C
Resolution 0.1°C
J 0.7°C
K 0.8°C
T 0.8°C
E 0.7°C
R 1.8°C
S 1.5°C
B 1.4°C
L 0.7°C
U 0.75°C
N 0.9°C
Resolution J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F
XK 0.6°C
BP 1.2°C
Source accuracy NI-120 0.2°C
PT-100 (385) 0.33°C
PT-100 (393) 0.3°C
PT-100 (JIS) 0.3°C
PT-200 (385) 0.2°C
PT-500 (385) 0.3°C
PT-1000 (385) 0.2°C
Resolution 0.1°C
Accuracy stated for 4-wire measurement.
J 0.7°C
K 0.8°C
T 0.8°C
E 0.7°C
R 1.4°C
S 1.5°C
B 1.4°C
L 0.7°C
U 0.75°C
N 0.9°C
Resolution J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C
XK 0.6°C
BP 1.2°C
Specifications
Ramp functions Source functions Voltage, current, resistance, frequency, temperature
Ramps Slow ramp, Fast ramp, 25% step-ramp
Loop power function Voltage 24 V
Accuracy 10%
Maximum current 22 mA, short circuit protected
Step functions Source functions Voltage, current, resistance, frequency, temperature
Steps 25% of range, 100% of range
Environmental Specifications
Operating temperature -10°C to +55°C
Storage temperature -20°C to 71°C
Operating altitude 3000 m
Safety Specifications
Agency approvals EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992
Mechanical and General Specifications
Size 130 x 236 x 61 mm
Weight 0.65 kg
Batteries 4 AA alkaline batteries
Battery replacement Separate battery compartment, accessable without breaking calibration seal
Side port connections Pressure module connector, also used for remote real-time programming

ضمانت‌نامه

این دستگاه دارای ضمانت‌نامه سه ساله شرکت امروز است.

متعلقات

این دستگاه شامل لوازم زیر است:

One set of AC72 alligator clips

One set of stackable test leads
One set of TL 75 testleads
User manuals
NIST- traceable calibration report and data

کاتالوگ

Fluke 725 Multifunction Process Calibrator